Model :
Slope :
# NH31501
56 + / - 3 mV/decade
Reproducibility :
+ / - 2%
Interferences :
Volatile amines
Temperature Range :
0 to 50¡Æ C (if calibrated at that temperature)
Pressure Range :
0 to 10 psi
ResponseTime :
95% response in 30 seconds
Storage :
Long term, store dry
Short term, store in ammonia standard with ISA added
Concentration Range :
5 x 10-7 M to 1M/ 0.01 ppm to 17,000 ppm NH3
pH Range :
Above 11 pH
Resistance :
1000 to 1500 megaohms
Care, Maintenance, & Cleaning :
Treat with a dilute ammonia standard with pH adjusted
> pH 11 Replace membrane as needed
Temperature Compensation :
No
Isopotential Point :
Unknown
On-Line Capability :
Yes, pH control required
Applications
High Purity Power Station Waters, Fish Tanks, Sea Waters, Waste Waters, Plating Baths, Biological Samples