Nitrogen Oxide Gas Sensing Electrode, Combination, Epoxy Body
Model :
Slope :
# NOX1502
56 + / - 3 mV / decade
Reproducibility :
+ / - 2%
Interferences :
CO2, SO2, HF, acetic acid
Temperature Range :
0 to 50¡Æ C (if calibrated at that temperature)
Pressure Range :
0 to 10 psi
ResponseTime :
95% response in 30 seconds
Storage :
Long term, store dryShort term, store in nitrogen oxide standard with ISA added
Concentration Range :
5 x 10-6 to 5 x 10-3 M0.2 ppm to 220 ppm
pH Range :
1.1 to 1.7 pH
Resistance :
1000 to 1500 megaohms
Care, Maintenance, & Cleaning :
Treat with a dilute nitrogen oxide standard with pH
adjusted to 1.1 - 1.7 Replace membrane as needed
Temperature Compensation :
No
Isopotential Point :
Unknown
On-Line Capability :
Yes, pH control required
Applications :
Air, Stack Gases