Model :
Slope :
# CO21501
55 + / - 3 mV/decade
Reproducibility :
+ / - 2%
Interferences :
Volatile weak acids
Complexation :
Heavy metals
Temperature Range :
0 to 50¡Æ C (if calibrated at that temperature)
Pressure Range :
0 to 10 psi
ResponseTime :
95% response in 30 seconds
Storage :
Long term, store dry Short term, store in carbon
dioxide standard with ISA added
Concentration Range :
10-4 to 10-2M 4.4 ppm to 440 ppm
pH Range :
4.8 to 5.2 pH
Resistance :
1000 to 1500 megaohms
Care, Maintenance, & Cleaning :
Treat with dilute carbon dioxide standard with pH
adjusted to 4.8 - 5.2 Replace membrane as needed
Temperature Compensation :
No
Isopotential Point :
Unknown
On-Line Capability :
Yes, pH control required
Applications :
Soft Drinks, Wine Fermentation
ISE Electrode
Calibration Standard solution
Filling solution
125ml
Standard solution 0.1 M
475ml
Standard solution 10 ppm
Standard solution 100 ppm
Standard solution 1,000 ppm
Standard solution 10,000 ppm
ISA (Ionic Strength Adjuster) solution
ISA solution
±âŸ
Storage solution
125ml
Membrane kit
Ä«Æ®¸®Áö-3°³, O-ring-3°³