Model | HC960 | HC960RS |
Measuring item | Conductivity
/ Salinity / Resistivity / Temp. |
Range | 0.000
uS/cm~200.0 mS/cm
0.0 ~ 70.0 ppt
0.00 ~ 20.00 M¥Ø.cm |
Resolution | 0.001
uS/cm
0.1 ppt
0.01 M¥Ø.cm |
Accuracy | ¡¾1% ¡¾1 digit |
Temp. | -30.0 ~ 130.0 ¡É |
Temp. accuracy | ¡¾0.2 ¡ÆC ¡¾1digit |
Temp. Compensation | Pt1000/NTC30K |
Remark | 2-cell, 4-cell both
electrodes available |
Electrode cell factor | choose
1 among 0.01, 0.1, 0.7, 10.00 1/cm
adjustable between 0.0050~19.99 1/cm |
Calibratin | 1 point |
Out
put | Analog output 1 : Isolated DC 0/4~20mA
corresponding to
measurement max. load
500¥Ø
Analog output 2 : Isolated DC
0/4~20mA corresponding to
Temperature max. load 500¥Ø |
Communication | x | RS485
(MODBUS) |
Contact
out | contact
output : Relay ON/OFF, max. 240VAC 0.5A
Notification : HI / LO Control point |
Washing
setting | ON : 0~9999 seconds / OFF :
0~999.9 hours |
Water
proof | IP 65 |
Temperature | 0 ~ 50 ¡É |
Dimension | 144 x 144 x 115 mm / 0.4 kg |
Power
supply | 100 ~ 240VAC / ¡¾10%,
50/60Hz |